FDA Adverse Event Injury Summary report: N

ACCOLADE PLUS TMZF HIP STEM #2

MDR report key: 3895163 · Received June 25, 2014

Report

Report Number
0002249697-2014-02433
Event Type
Injury
Date Received
June 25, 2014
Date of Event
June 3, 2014
Report Date
June 3, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MEH
PMA / PMN Number
K994366
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF RECEIVED, WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT UPON COMPLETION OF THE INVESTIGATION. DEVICE REMAINS IMPLANTED.

Additional Manufacturer Narrative · 1

AN EVENT REGARDING PSEUDOTUMOR AND CORROSION INVOLVING AN ACCOLADE PLUS TMZF HIP STEM #2 WAS REPORTED. THE EVENT WAS NOT CONFIRMED. DEVICE EVALUATION NOT PERFORMED AS NO DEVICE WAS RETURNED. A REVIEW OF THE PROVIDED MEDICAL RECORDS BY A CLINICAL CONSULTANT INDICATED NO CERTAIN ROOT CAUSE OF FAILURE CAN BE ESTABLISHED FOR THIS CASE DUE TO LACK OF PROPER INFORMATION. A DEVICE HISTORY REVIEW CONFIRMED ALL DEVICES ACCEPTED INTO FINISHED GOODS CONFORMED TO SPECIFICATION. A COMPLAINT HISTORY REVIEW CONFIRMED NO OTHER SIMILAR EVENTS FOR THE REPORTED LOT. THE EXACT CAUSE OF THE EVENT COULD NOT BE DETERMINED BECAUSE OF A LACK OF INFORMATION. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME.

Description of Event or Problem · 1

PATIENT PRESENTED WITH GROIN PAIN, APPROXIMATELY 6 YEARS AFTER HIP IMPLANTED. SCANS SHOWED PSEUDOTUMOR. REVISION OF RIGHT HIP, POSSIBLE PSEUDO TUMOR, TRUNNION SHOWED SIGNS OF POSSIBLE CORROSION. PSEUDOTUMOR EXCISED AND FEMORAL HEAD CHANGED. SURGEON CHOSE TO REMOVE COCR FEMORAL HEAD AND REPLACE WITH CERAMIC WITHOUT A SLEEVE. SURGEON IS AWARE THAT NOT USING A SLEEVE IS OFF LABEL.

Description of Event or Problem · 1

PATIENT PRESENTED WITH GROIN PAIN, APPROXIMATELY 6 YEARS AFTER HIP IMPLANTED. SCANS SHOWED PSEUDOTUMOR. REVISION OF RIGHT HIP, POSSIBLE PSEUDO TUMOR, TRUNNION SHOWED SIGNS OF POSSIBLE CORROSION. PSEUDOTUMOR EXCISED AND FEMORAL HEAD CHANGED. SURGEON CHOSE TO REMOVE COCR FEMORAL HEAD AND REPLACE WITH CERAMIC WITHOUT A SLEEVE. SURGEON IS AWARE THAT NOT USING A SLEEVE IS OFF LABEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
372163 ACCOLADE PLUS TMZF HIP STEM #2 IMPLANT MEH STRYKER ORTHOPAEDICS-MAHWAH 23492302

Patients

Seq Age Sex Outcome Treatment
1 71 YR Hospitalization| R