STAPLE, IMPLANTABLE
Report
- Report Number
- 3005075853-2014-04397
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Date of Event
- March 20, 2014
- Report Date
- May 6, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
(B)(4) = DAMAGED FIRING TRIGGER TEETH. SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AND WITH A 6R45B RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED FULLY LOADED WITH STAPLES WITH THE LOCKOUT SPRING NORMAL. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. WHILE NO CONCLUSION COULD BE REACHED ON WHAT CAUSED THE FIRING MECHANISM TO FAIL, IT IS POSSIBLE THAT THE DEVICE WAS ATTEMPTED TO FIRE ON TICKER TISSUE THAN INDICATED OR ATTEMPTED TO FIRE TROUGH A LOCKED RELOAD IN PREVIOUS FIRINGS CAUSING AN INCREASE OF THE INTERNAL FORCES RESULTING IN THE COMPONENT YIELDING. IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4).
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTROPLASTY PROCEDURE, WHEN TRIGGERED THE GAUNTLET FOR STAPLING, THE STAPLER BROKE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 371531 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | K4DP14 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |