FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 3894904 · Received June 25, 2014

Report

Report Number
2955842-2014-03897
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
June 2, 2014
Report Date
June 2, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K061260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND A BROKEN PITCH UP CABLE AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SURGICAL PROCEDURE, A BROKEN CABLE WAS IDENTIFIED ON A PK DISSECTING FORCEPS INSTRUMENT. THERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT. NOTHING REPORTEDLY FELL INTO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
370386 PK DISSECTING FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420227-03 N10140409 223

Patients

Seq Age Sex Outcome Treatment
1