LEAD MODEL 304
Report
- Report Number
- 1644487-2014-01589
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Date of Event
- May 30, 2014
- Report Date
- May 30, 2014
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE MANUFACTURING RECORDS WERE REVIEWED. REVIEW OF MANUFACTURING RECORDS CONFIRMED THAT THE LEAD PASSED ALL FUNCTIONAL TESTS PRIOR TO DISTRIBUTION. DEVICE FAILURE IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
ONLY A PORTION OF THE LEAD WAS RETURNED FOR ANALYSIS WHICH DID NOT REVEAL ANY ANOMALIES. DEVICE FAILURE IS SUSPECTED IN THE LEAD PORTION NOT RETURNED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
THE EXPLANTED GENERATOR AND LEAD WERE RECEIVED BY THE MANUFACTURER ON 12/17/2014. HOWEVER, ANALYSIS HAS NOT BEEN COMPLETED TO DATE.
IT WAS REPORTED THAT THE VNS PATIENT¿S DEVICE SHOWED HIGH IMPEDANCE. THE PATIENT WAS SENT FOR X-RAYS. NO KNOWN SURGICAL INTERVENTIONS HAVE OCCURRED TO DATE. ATTEMPTS FOR ADDITIONAL RELEVANT INFORMATION HAVE BEEN UNSUCCESSFUL TO DATE.
ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE VNS PATIENT UNDERWENT LEAD REPLACEMENT SURGERY ON (B)(6) 2014. THE EXPLANTING FACILITY WILL NOT RETURN EXPLANTED DEVICES TO THE MANUFACTURER FOR ANALYSIS; THEREFORE, NO ANALYSIS CAN BE PERFORMED.
ANALYSIS OF THE EXPLANTED DEVICES WAS COMPLETED. THERE WERE NO PERFORMANCE OR ANY OTHER TYPE OF ADVERSE CONDITIONS FOUND WITH THE PULSE GENERATOR THAT OCCURRED DURING IMPLANT. NOTE THAT A PORTION OF THE LEAD ASSEMBLY (BODY) INCLUDING THE ELECTRODES WAS NOT RETURNED FOR ANALYSIS AND THEREFORE A COMPLETE EVALUATION COULD NOT BE PERFORMED ON THE ENTIRE LEAD PRODUCT. THE ABRADED OPENING AND TOOL MARKS FOUND ON THE OUTER SILICONE TUBING AND THE CUT ENDS THAT WERE MADE DURING THE EXPLANTED PROCESS, MOST LIKELY PROVIDED THE LEAKAGE PATH FOR THE DRIED REMNANTS OF WHAT APPEARED TO HAVE ONCE BEEN BODY FLUIDS FOUND INSIDE THE OUTER SILICONE TUBING. FOR THE OBSERVED INNER TUBING FLUID REMNANTS, THERE WAS NO OBVIOUS PATH FOR FLUID INGRESS OTHER THAN THE CUT ENDS THAT WERE MADE DURING THE EXPLANTED PROCESS. THE CONDITION OF THE RETURNED LEAD PORTION IS CONSISTENT WITH CONDITIONS THAT TYPICALLY EXIST FOLLOWING AN EXPLANT PROCEDURE. NO OBVIOUS ANOMALIES WERE NOTED. THE SETSCREW MARKS FOUND ON THE LEAD CONNECTOR PINS PROVIDE EVIDENCE THAT, AT ONE POINT IN TIME, A GOOD MECHANICAL AND ELECTRICAL CONNECTION WAS PRESENT. CONTINUITY CHECKS OF THE RETURNED LEAD PORTION WERE PERFORMED, DURING THE VISUAL ANALYSIS, AND NO DISCONTINUITIES WERE IDENTIFIED. BASED ON THE FINDINGS IN THE PRODUCT ANALYSIS LAB, THERE IS NO EVIDENCE TO SUGGEST DISCONTINUITIES IN THE RETURNED PORTION OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 370141 | LEAD MODEL 304 | LEAD | LYJ | CYBERONICS, INC. | 304-20 | 201852 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR |