PDS PLUS ANTIBACTERIAL (POLYDIOXANONE) SUTURE
Report
- Report Number
- 2210968-2014-08207
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Date of Event
- June 7, 2014
- Report Date
- June 9, 2014
- Manufacturer
- ETHICON INC.
- Product Code
- NEW
- PMA / PMN Number
- K061037
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
CONCLUSION: ACTUAL DEVICE WAS RETURNED FOR EVALUATION. MICROSCOPIC VISUAL EXAMINATION ON NEEDLE THAT WAS BROKEN AT THE ATTACHMENT END REVEALED INDENTS AND SCUFF MARKS AROUND THE BREAK AREA PRODUCED DURING HANDLING BY A SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVICE. THE NEEDLE FRACTURED DUE TO TENSILE OVERLOAD GENERATED DURING SEVERE MECHANICAL DEFORMATION, WITH SIGNS OF DUCTILE FAILURE. THERE ARE NO SIGNS OF MANUFACTURING DEFECTS FOUND ON THE NEEDLE. THE DEVICE INFORMATION FOR USE CAUTIONS THE USER THAT "TO AVOID DAMAGING NEEDLE POINTS AND SWAGE AREAS, GRASP THE NEEDLE IN AN AREA ONE-THIRD (1/3) TO ONE-HALF (1/2) OF THE DISTANCE FROM THE SWAGED END TO THE POINT." ADDITIONAL INFORMATION: THE ACTUAL DEVICE BATCH NUMBER ASSOCIATED WITH THIS EVENT IS NOT KNOWN. TWO POSSIBLE BATCH NUMBERS ARE REPORTED AS FOLLOWS: BATCH HB2422, BATCH HC2025. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS FOR THE POSSIBLE BATCH NUMBER HB2422, WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.
(B)(4). CONCLUSION: TO DATE, THE DEVICE HAS NOT BEEN RETURNED. IF THE PRODUCT IS RETURNED FOR EVALUATION, ANY FURTHER INFO DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
IT WAS REPORTED THAT A PATIENT UNDERWENT A CESAREAN SECTION ON (B)(6) 2014 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE WHEN THE SURGEON WAS MAKING A KNOTTED SUTURE THROUGH THE DERMIS. THE SURGEON OPINED THAT THE NEEDLE MIGHT HAVE BROKE DUE TO EXTRA FORCE APPLIED TO IT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 370281 | PDS PLUS ANTIBACTERIAL (POLYDIOXANONE) SUTURE | SUTURE, ABSORBABLE | NEW | ETHICON INC. | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |