ROTATING MULTIPLE CLIP APPLIER
Report
- Report Number
- 3005075853-2014-04385
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Date of Event
- May 20, 2014
- Report Date
- May 22, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
(B)(4). YIELDED JAW. THE ER320 DEVICE WAS RETURNED FOR ANALYSIS AND UPON INSPECTION THE JAWS WERE FOUND TO BE IN A YIELDED CONDITION AND WITH A CLIP IN JAWS. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING THE DEVICE FOR FUNCTIONALITY THE DEVICE WAS CYCLED, FED, AND FORMED 8 CLIPS AS INTENDED. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. HOWEVER, DURING TESTING NO DROPPING OR EJECTED CLIPS WERE NOTED. IT IS KNOWN FROM THE HISTORY OF THE DEVICE THAT THE CONDITION OF THE JAWS MAY LEAD DROPPING/EJECTED CLIPS. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE CLIPS KEEP FALLING OUT OF THE JAW OF THE DEVICE WHEN A LITTLE TENSION IS ON IT. WHEN TISSUE IS MOVED A LITTLE WITH THE JAW, THE CLIP IMMEDIATELY FALLS OUT. THE PROCEDURE WAS FINISHED WITHOUT THE USE OF A CLIP APPLIER. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 369901 | ROTATING MULTIPLE CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | L4ED3P |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |