FDA Adverse Event Malfunction Summary report: N

SYMBIQ SINGLE CHANNE

MDR report key: 3894588 · Received May 21, 2014

Report

Report Number
9615050-2014-03528
Event Type
Malfunction
Date Received
May 21, 2014
Date of Event
January 1, 2014
Report Date
April 21, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING THE DEVICE ALARMED WITH A S233 MALFUNCTION ALARM CODE AND WAS NOTED IN THE DEVICE HISTORY. IT WAS NOTED THAT THE DEVICE FAN WAS NOT WORKING. THE PROBABLE CAUSE WAS DUE TO A BROKEN DEVICE COOLING FAN. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE DEVICE ALARMED WITH A S233 (OVER TEMPERATURE ERROR) MALFUNCTION ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN USE. DURING TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH A S233 MALFUNCTION ALARM CODE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303938 SYMBIQ SINGLE CHANNE 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK