FDA Adverse Event Malfunction Summary report: N

EMAX 2 PLUS MOTOR

MDR report key: 3894573 · Received June 25, 2014

Report

Report Number
1045834-2014-12295
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
June 9, 2014
Report Date
June 10, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK080802
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. DURING TESTING THE DEVICE WAS FOUND TO HAVE LOW RPM AND THE AIR PUMP STAYS ON WHICH CAUSES A SLIGHT VIBRATION IN THE HOSE. THE REPORTED CONDITIONS WERE CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE THE MOTOR DEVICE OPERATED FOR LESS THAN ONE MINUTE, THEN SLOWED AND STOPPED IN THE SPACE OF A FEW SECONDS. A CODE E6 WAS DISPLAYED ON THE CONSOLE DEVICE. ACCORDING TO THE REPORT, EVEN AFTER THE PEDAL WAS NO LONGER DEPRESSED, THE WIRING THAT WAS EXITING THE HAND PIECE COULD BE FELT TO BE VIBRATING. THE REPORTER STATED THAT THE ONLY WAY TO STOP THE VIBRATION WAS BY UNPLUGGING THE HAND PIECE FROM THE CART. THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE FOR USE. THE TYPE OF SPARE WAS NOT PROVIDED. THERE WAS PATIENT INVOLVEMENT REPORTED, PATIENT INFORMATION WAS NOT PROVIDED. THERE WERE NO PATIENT OR USER INJURIES REPORTED. IT WAS REPORTED THERE WAS NO MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
371653 EMAX 2 PLUS MOTOR MOTOR, DRILL, ELECTRIC - HANDPIECE HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1 CONSOLE DEVICE