ENDOTRACHEAL TUBES - AIRWAY MANAGEMENT
Report
- Report Number
- 9611710-2014-00082
- Event Type
- Malfunction
- Date Received
- May 2, 2014
- Date of Event
- April 7, 2014
- Report Date
- April 7, 2014
- Manufacturer
- UNOMEDICAL SDN BHD
- Product Code
- BTR
- PMA / PMN Number
- K951696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
AN EVALUATION WAS PERFORMED ON (B)(4) 2014. RECEIVED 24 UNUSED I.D. 6.0MM, CATALOG NUMBER: 61110060 ENDOTRACHEAL TUBES WITH FROSTY BALLOONS, FITTED WITH PP CONNECTORS OF CORRESPONDING SIZES AND TUBES MARKED AS PER JSS SPECIFICATION. PRODUCTS FITTED WITH PVC BULLET VALVES WITH PILOT BALLOONS PRINTED WITH SIZES IDENTIFICATION 6.0 MM AND WORK ORDER: NUMBER (B)(4). THE SAMPLES WERE TESTED ON (B)(4) 2014. THE SAMPLES WERE INSPECTED AND THE OBSERVATIONS ARE AS FOLLOWS: (B)(4) UNITS WERE CONFIRMED FOR STAIN ON BALLOON AND (B)(4) UNITS WERE CONFIRMED FOR STAIN IN PILOT BALLOON. REVIEW OF THE ASSEMBLY RECORD DID NOT REVEAL ANY SIGNS OF THE ASSOCIATED DEFECTS DETECTED DURING THE 100% LIGHT TEST. BASED ON THE ANALYSIS OF THE RETURNED SAMPLES PROVIDED IT WAS FOUND THAT THEY DID NOT MEET OUR INSPECTION CRITERIA. HOWEVER, THIS COMPLAINT ISSUE HAS BEEN INVESTIGATED TO ADDRESS THE CONFIRMED COMPLAINT AND DOCUMENTED IN THE CAPA SYSTEM. NO ADDITIONAL PATIENT/EVENT DETAILS HAVE BEEN PROVIDED TO DATE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW-UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT DURING INCOMING INSPECTION "FOUND STAIN ON PRODUCT."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 266130 | ENDOTRACHEAL TUBES - AIRWAY MANAGEMENT | TUBE, TRACHEAL (W/WO CONNECTOR) | BTR | UNOMEDICAL SDN BHD | 61110060 | 615275R001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |