FDA Adverse Event Malfunction Summary report: N

GEMSTR 7 THPY ALT ST

MDR report key: 3894486 · Received April 25, 2014

Report

Report Number
9615050-2014-02980
Event Type
Malfunction
Date Received
April 25, 2014
Date of Event
March 18, 2014
Report Date
March 19, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
Z-1169-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING THE DEVICE WOULD NOT POWER UP ON AC OR DC POWER. FURTHER TESTING FOUND LEAKAGE FROM THE DISPOSABLE BATTERIES IN THE BATTERY COMPARTMENT AND THE MIDDLE PRINTED CIRCUIT BOARD OF THE DEVICE. THE PROBABLY CAUSE OF NO AC OR DC POWER WAS THE CORROSION FROM THE BATTERY LEAKAGE ON THE MIDDLE PRINTED CIRCUIT BOARD. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT DURING ROUTINE TESTING AT THE USER FACILITY, THE DEVICE HAS " NO AC POWER". NO ADDITIONAL INFO WAS PROVIDED. THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER LEAKAGE WAS NOTED FROM THE DISPOSABLE BATTERIES IN THE BATTERY COMPARTMENT OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252705 GEMSTR 7 THPY ALT ST 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA