FDA Adverse Event Injury Summary report: N

HP MBT CEM KEEL PUNCH SZ 4-7

MDR report key: 3894478 · Received June 25, 2014

Report

Report Number
1818910-2014-21945
Event Type
Injury
Date Received
June 25, 2014
Date of Event
June 6, 2014
Report Date
June 6, 2014
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LXH
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE RETURNED DEVICE CONFIRMS THE REPORTED EVENT OF TAB BREAKAGE. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS ABOUT THE ROOT CAUSE OF THE FRACTURES, HOWEVER HEAVY USAGE OVER TIME AND OR USER ERROR/MISUSE COULD BE CONTRIBUTING FACTORS. (B)(4) WAS PREVIOUSLY INITIATED TO IDENTIFY ROOT CAUSE AND ANY CORRECTIVE ACTIONS. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE TABS ON THE TOP OF THE SIGMA HP MBT KEEL PUNCH SZ 4-7 BROKE OFF DURING USE. THE BROKEN PIECES WERE RECOVERED CHECKED AGAINST THE PUNCH TO ENSURE NO PIECES WERE LEFT UNACCOUNTED FOR AND THEN THE BROKEN TAB WAS DISCARDED. THE SIGMA HP MBT KEEL PUNCH SZ 4-7 HAS BEEN DECONTAMINATED AND WILL BE RETURNED. PLEASE NOTE THAT THE BROKEN TAB WAS RECOVERED AND CHECKED AGAINST THE PUNCH TO ENSURE ALL PIECES WERE ACCOUNTED FOR TO ENSURE THERE WAS NO THIRD BODY METAL MATERIAL LEFT IN SITU. THE WOUND WAS THEN EXTENSIVELY IRRIGATED AND SUCTIONED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
371149 HP MBT CEM KEEL PUNCH SZ 4-7 KNEE INSTRUMENT/TRIAL LXH DEPUY ORTHOPAEDICS, INC. J1108

Patients

Seq Age Sex Outcome Treatment
1 Other