FDA Adverse Event Malfunction Summary report: N

GEMSTAR PAIN MGE SNG

MDR report key: 3894475 · Received April 25, 2014

Report

Report Number
9615050-2014-02966
Event Type
Malfunction
Date Received
April 25, 2014
Date of Event
March 26, 2014
Report Date
March 26, 2014
Manufacturer
HOSPIRA HOLDINGS DE COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
Z-1169-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, LEAKAGE FROM THE DISPOSABLE BATTERIES WAS FOUND IN THE BATTERY COMPARTMENT AND THE PRINTED CIRCUIT BOARD OF THE DEVICE. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT TO REPLACE THE LITHIUM BATTERY. THIS IS NOT A REPORTABLE MALFUNCTION. DURING VERIFICATION TESTING AT THE SERVICE CENTER THE DEVICE WAS FOUND TO HAVE CORROSION IN THE BATTERY COMPARTMENT DUE TO LEAKAGE FROM THE DISPOSABLE BATTERIES.