GEMSTAR PAIN MGE SNG
MDR report key: 3894475
·
Received April 25, 2014
Report
- Report Number
- 9615050-2014-02966
- Event Type
- Malfunction
- Date Received
- April 25, 2014
- Date of Event
- March 26, 2014
- Report Date
- March 26, 2014
- Manufacturer
- HOSPIRA HOLDINGS DE COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K060806
- Removal / Correction Number
- Z-1169-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING TESTING, LEAKAGE FROM THE DISPOSABLE BATTERIES WAS FOUND IN THE BATTERY COMPARTMENT AND THE PRINTED CIRCUIT BOARD OF THE DEVICE. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT TO REPLACE THE LITHIUM BATTERY. THIS IS NOT A REPORTABLE MALFUNCTION. DURING VERIFICATION TESTING AT THE SERVICE CENTER THE DEVICE WAS FOUND TO HAVE CORROSION IN THE BATTERY COMPARTMENT DUE TO LEAKAGE FROM THE DISPOSABLE BATTERIES.