SYMBIQ SINGLE CHANNE

MDR report key: 3894470 · Received April 25, 2014

Report

Report Number
9615050-2014-02970
Event Type
Malfunction
Date Received
April 25, 2014
Date of Event
January 1, 2014
Report Date
March 26, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0069-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE REP PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, THE DEVICE TOUCHSCREEN WAS FOUND TO NOT RESPOND WHEN PRESSED. THIS WAS DUE TO THE TOUCHSCREEN BEING OUT OF CALIBRATION. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE DEVICE TOUCHSCREEN DOES NOT RESPOND WHEN PRESSED. THE DEVICE WAS RETURNED TO THE BIOMED DEPT FOR AN UNSPECIFIED REASON. NO INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING THE TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252650 SYMBIQ SINGLE CHANNE 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK