SYMBIQ DCHANNEL 3.01
Report
- Report Number
- 9615050-2014-02989
- Event Type
- Malfunction
- Date Received
- April 25, 2014
- Date of Event
- March 1, 2014
- Report Date
- March 28, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z-0070-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
TESTING AND INVESTIGATION FOUND THE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. FURTHER TESTING FOUND THE DEVICE TOUCHSCREEN KEYS WOULD RANDOMLY GET STUCK. THE PROBABLE CAUSE WAS DUE TO CONTAMINATION ON THE BOTTOM OF THE TOUCHSCREEN DUE TO FLUID INGRESS. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN NOT WORKING. THE DEVICE WAS RETURNED TO THE BIOMED DEPT WITH A REPORT OF DEVICE TOUCHSCREEN NOT WORKING.. NO INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY,THE DEVICE DID NOT PASS THE TOUCHSCREEN TEST. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 252181 | SYMBIQ DCHANNEL 3.01 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |