FDA Adverse Event Malfunction Summary report: N

SYMBIQ DCHANNEL 3.01

MDR report key: 3894438 · Received April 25, 2014

Report

Report Number
9615050-2014-02988
Event Type
Malfunction
Date Received
April 25, 2014
Date of Event
March 1, 2014
Report Date
March 28, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0070-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THAT THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED AND DID NOT PRODUCE AN AUDIBLE TOUCH TONE. THE PROBABLE CAUSE WAS DUE TO CONTAMINATION ON THE TOUCHSCREEN DUE TO FLUID INGRESS. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN NOT WORKING. THE DEVICE WAS RETURNED TO THE BIOMED DEPT WITH A REPORT OF DEVICE TOUCHSCREEN NOT WORKING. NO INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT PASS THE TOUCHSCREEN TEST. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252345 SYMBIQ DCHANNEL 3.01 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK