FDA Adverse Event Malfunction Summary report: N

GEMSTAR PM PMP

MDR report key: 3894430 · Received April 25, 2014

Report

Report Number
9615050-2014-03008
Event Type
Malfunction
Date Received
April 25, 2014
Date of Event
February 15, 2014
Report Date
March 6, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K083019
Removal / Correction Number
Z-1170-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, LEAKAGE WAS NOTED FROM THE DISPOSABLE BATTERIES IN THE BATTERY COMPARTMENT, TOP AND MIDDLE PRINTED CIRCUIT BOARD AND BATTERY CAP OF THE DEVICE. THE CUSTOMER REPORTED 10/000/000 WAS DUPLICATED DURING TESTING. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH A 18/000 (HISTORY POINTER ERROR WHILE INSERTING A NEW RECORD INTO HISTORY LOG) MALFUNCTION ALARM CODE. NO ADDITIONAL INFO WAS PROVIDED. THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, BATTERY LEAKAGE WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252153 GEMSTAR PM PMP 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA