GEMSTAR PM PMP
Report
- Report Number
- 9615050-2014-03008
- Event Type
- Malfunction
- Date Received
- April 25, 2014
- Date of Event
- February 15, 2014
- Report Date
- March 6, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K083019
- Removal / Correction Number
- Z-1170-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
DURING TESTING, LEAKAGE WAS NOTED FROM THE DISPOSABLE BATTERIES IN THE BATTERY COMPARTMENT, TOP AND MIDDLE PRINTED CIRCUIT BOARD AND BATTERY CAP OF THE DEVICE. THE CUSTOMER REPORTED 10/000/000 WAS DUPLICATED DURING TESTING. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH A 18/000 (HISTORY POINTER ERROR WHILE INSERTING A NEW RECORD INTO HISTORY LOG) MALFUNCTION ALARM CODE. NO ADDITIONAL INFO WAS PROVIDED. THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, BATTERY LEAKAGE WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 252153 | GEMSTAR PM PMP | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |