COULTER LH 750 HEMATOLOGY ANALYZER

MDR report key: 3894411 · Received June 25, 2014

Report

Report Number
1061932-2014-01467
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
June 3, 2014
Report Date
June 3, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K011342
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2014, THE BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT AND FOUND CUT TUBING THROUGH PINCH VALVE VL53B. THE FSE REPLACED THE TUBING AND THE INSTRUMENT RAN WITHOUT ANY LEAKS. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED A LEAK WHEN USING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE VOLUME OF THE LEAK WAS ABOUT 2 ML AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER COULD NOT IDENTIFY THE SOURCE OF THE LEAK. THE OPERATOR WAS WEARING A LAB COAT AND GLOVES AT THE TIME OF THE INCIDENT. THERE WAS NO BIOHAZARD EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. THERE WAS NO REPORT OF ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.