FDA Adverse Event Malfunction Summary report: N

PLUM A+ DRIVER ED 2

MDR report key: 3894369 · Received April 23, 2014

Report

Report Number
9615050-2014-02918
Event Type
Malfunction
Date Received
April 23, 2014
Date of Event
February 13, 2014
Report Date
February 13, 2014
Manufacturer
HOSPIRA COSTA RICA LTD
Product Code
FRN
PMA / PMN Number
K042081
Removal / Correction Number
Z-0624-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS MISSING AND THE DOOR ROLLER PIN WAS BROKEN. FURTHER TESTING FOUND THE DEVICE HAD UNRESTRICTED FLOW WHEN THE DOOR WAS OPENED. THIS WAS DUE TO THE MISSING DEVICE DOOR ROLLER AND BROKEN DOOR ROLLER PIN. THE MISSING DEVICE DOOR ROLLER AND BROKEN DOOR ROLLER PIN PREVENTED PROPER ACTUATION OF THE REGULATOR CLOSER ON THE DEVICE MECHANISM CAUSING UNRESTRICTED FLOW WHEN THE DOOR WAS OPENED. THE CUSTOMER CONTACT'S REPORTED BROKEN CE MODULE ANTENNA HOUSING WAS DUPLICATED DURING TESTING. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT THE CE MODULE ANTENNA HOUSING WAS SMASHED. THIS DOES NOT INDICATE A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS MISSING AND THE DOOR ROLLER PIN WAS BROKEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
245238 PLUM A+ DRIVER ED 2 80FRN FRN HOSPIRA COSTA RICA LTD NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA PLUM A+ SOFTWARE MODULE, LIST #20791,| (B)(4)