PLUM A+ DRIVER ED 2
Report
- Report Number
- 9615050-2014-02918
- Event Type
- Malfunction
- Date Received
- April 23, 2014
- Date of Event
- February 13, 2014
- Report Date
- February 13, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD
- Product Code
- FRN
- PMA / PMN Number
- K042081
- Removal / Correction Number
- Z-0624-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS MISSING AND THE DOOR ROLLER PIN WAS BROKEN. FURTHER TESTING FOUND THE DEVICE HAD UNRESTRICTED FLOW WHEN THE DOOR WAS OPENED. THIS WAS DUE TO THE MISSING DEVICE DOOR ROLLER AND BROKEN DOOR ROLLER PIN. THE MISSING DEVICE DOOR ROLLER AND BROKEN DOOR ROLLER PIN PREVENTED PROPER ACTUATION OF THE REGULATOR CLOSER ON THE DEVICE MECHANISM CAUSING UNRESTRICTED FLOW WHEN THE DOOR WAS OPENED. THE CUSTOMER CONTACT'S REPORTED BROKEN CE MODULE ANTENNA HOUSING WAS DUPLICATED DURING TESTING. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT THE CE MODULE ANTENNA HOUSING WAS SMASHED. THIS DOES NOT INDICATE A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS MISSING AND THE DOOR ROLLER PIN WAS BROKEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 245238 | PLUM A+ DRIVER ED 2 | 80FRN | FRN | HOSPIRA COSTA RICA LTD | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | PLUM A+ SOFTWARE MODULE, LIST #20791,| (B)(4) |