FDA Adverse Event Malfunction Summary report: N

COULTER LH 780 HEMATOLOGY ANALYZER

MDR report key: 3894353 · Received June 25, 2014

Report

Report Number
1061932-2014-01453
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
June 3, 2014
Report Date
June 3, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061616
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2014, A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE ANALYZER AND DID NOT OBSERVE A LEAK OR ERROR AND WAS UNABLE TO REPLICATE THE LEAK AFTER STARTUP. THE FSE IDENTIFIED THAT THE INSTRUMENT WAS GENERATING AN ERROR CONDITION OF VOTEOUTS (NON-NUMERIC RESULTS) FOR THE DIFFERENTIAL PARAMETERS. THE FSE DETERMINED THE DIFF (DIFFERENTIAL) MIXING MOTOR WAS NOT MIXING PROPERLY BECAUSE A CABLE WAS LOOSE. THE FSE RECONNECTED THE DIFF MIXING MOTOR AND THE INSTRUMENT RAN WITHOUT VOTEOUTS ON THE DIFFERENTIAL PARAMETERS. THE INSTRUMENT WAS RUNNING WITHOUT LEAKS OR ERRORS FOLLOWING THE SERVICE. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED A LEAK WHEN USING THE COULTER LH 780 HEMATOLOGY ANALYZER. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER'S FACILITY TO EVALUATE THE ANALYZER. THE FSE COULD NOT CONFIRM THE LEAK. THE FSE DID OBSERVE THE INSTRUMENT GENERATED AN ERROR CONDITION OF VOTEOUTS (NON-NUMERIC RESULTS) FOR THE DIFFERENTIAL PARAMETERS. THE FSE DETERMINED THE CAUSE OF THE VOTEOUTS TO BE A LOOSE CABLE ON THE DIFFERENTIAL MIXING MOTOR. THE OPERATOR WAS WEARING A LAB COAT AND GLOVES AT THE TIME OF THE INCIDENT. THERE WAS NO REPORT OF BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THERE WAS NO REPORT OF ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO PATIENT TREATMENT.