FDA Adverse Event Malfunction Summary report: N

COULTER LH 780 HEMATOLOGY ANALYZER

MDR report key: 3894350 · Received June 25, 2014

Report

Report Number
1061932-2014-01451
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
June 3, 2014
Report Date
June 3, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061616
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2014, THE CUSTOMER IDENTIFIED THE SOURCE OF THE LEAK WAS TUBING THAT WAS DISCONNECTED FROM THE BLOOD SAMPLING VALVE (BSV). THE CUSTOMER REATTACHED THE TUBING AND THE INSTRUMENT RAN WITHOUT LEAKS. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE ANALYZER AND VERIFIED THAT THE CUSTOMER HAD FIRMLY REATTACHED THE TUBING. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED A LEAK WHEN USING THE COULTER LH 780 HEMATOLOGY ANALYZER. THE LEAK WAS FROM UNDER THE BLOOD SAMPLING VALVE (BSV) INSIDE THE INSTRUMENT. THE VOLUME OF THE LEAK WAS ABOUT 5 ML AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING A LAB COAT, GOGGLES, AND GLOVES AT THE TIME OF THE INCIDENT. THERE WAS NO REPORT OF BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THERE WAS NO REPORT OF ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
370668 COULTER LH 780 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1