GEMSTAR PAIN MGE SNG
Report
- Report Number
- 9615050-2014-02910
- Event Type
- Malfunction
- Date Received
- April 23, 2014
- Date of Event
- March 10, 2014
- Report Date
- March 11, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD
- Product Code
- FRN
- PMA / PMN Number
- K060806
- Removal / Correction Number
- Z-1169-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
DURING TESTING LEAKAGE FORM THE DISPOSABLE BATTERIES WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE. THE CUSTOMER REPORTED CHECK CASSETTE ALARM WAS NOT DUPLICATED; HOWEVER, WAS NOTED IN THE DEVICE HISTORY. THIS WAS DUE TO A BROKEN CASSETTE LATCH. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT DURING ROUTINE TESTING, THE DEVICE ALARMED WITH A CHECK CASSETTE ALARM. THIS DOES NOT INDICATE A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER LEAKAGE FORM THE DISPOSABLE BATTERIES WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 245871 | GEMSTAR PAIN MGE SNG | 80FRN | FRN | HOSPIRA COSTA RICA LTD | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |