FDA Adverse Event Malfunction Summary report: N

GEMSTAR PAIN MGE SNG

MDR report key: 3894343 · Received April 23, 2014

Report

Report Number
9615050-2014-02910
Event Type
Malfunction
Date Received
April 23, 2014
Date of Event
March 10, 2014
Report Date
March 11, 2014
Manufacturer
HOSPIRA COSTA RICA LTD
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
Z-1169-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING LEAKAGE FORM THE DISPOSABLE BATTERIES WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE. THE CUSTOMER REPORTED CHECK CASSETTE ALARM WAS NOT DUPLICATED; HOWEVER, WAS NOTED IN THE DEVICE HISTORY. THIS WAS DUE TO A BROKEN CASSETTE LATCH. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT DURING ROUTINE TESTING, THE DEVICE ALARMED WITH A CHECK CASSETTE ALARM. THIS DOES NOT INDICATE A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER LEAKAGE FORM THE DISPOSABLE BATTERIES WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
245871 GEMSTAR PAIN MGE SNG 80FRN FRN HOSPIRA COSTA RICA LTD NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA