FDA Adverse Event Malfunction Summary report: N

PLUM A+ DRIVER ED 2

MDR report key: 3894330 · Received April 23, 2014

Report

Report Number
9615050-2014-02902
Event Type
Malfunction
Date Received
April 23, 2014
Report Date
May 22, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K042081
Removal / Correction Number
Z-0624-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. THE CUSTOMER CONTACT'S REPORTED MALFUNCTION 430 (PROXIMAL AIR SENSOR FAIL) WAS NOT DUPLICATED DURING TESTING. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT THE DEVICE ALARMED WITH AN 430 (PROXIMAL AIR SENSOR FAIL). THIS DOES NOT INDICATE A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
245779 PLUM A+ DRIVER ED 2 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA