ZIMMER SKIN GRAFT MESHER
Report
- Report Number
- 1526350-2014-00094
- Event Type
- Malfunction
- Date Received
- February 14, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 17, 2014
- Manufacturer
- ZIMMER SURGICAL
- Product Code
- GFD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- UNKNOWN
Narratives
THE DEVICE WAS RETURNED TO THE MFR FOR REPAIR AND EVAL. THE SERVICE RECORD INDICATES THAT THE DEVICE WAS MANUFACTURED ON 09/08/2010 AND WAS LAST REPAIRED ON 06/19/2013 FOR A NON-RELATED ISSUE. EVAL OF THE DEVICE OBSERVED WORN GEARS, EXCESSIVE GALLING OF THE ROLLER GEAR AND CUTTER GEAR TEETH. MINOR WEAR TO THE SHOULDER BOLTS AND DAMAGE TO THE SIDE PLATES AND ROLLER WERE ALSO OBSERVED. PRIOR TO REPAIR, A TEST MESH AND CALIBRATION CHECK COULD NOT BE PERFORMED DUE TO THE DAMAGED GEARS, THE CUSTOMER RETURNED CUTTERS FOR EVAL. EXCESSIVE GALLING OF THE CUTTER GEAR TEETH WAS OBSERVED FOR BOTH CUTTERS. THE GEARS AND COLLARS WERE REPLACED TO BOTH CUTTERS, WHICH PASSED CUTTER EVAL. IT WAS NOTED THAT THE CUSTOMER IS A CADAVER TISSUE BANK WHICH EXPERIENCES HEAVY USAGE OF DEVICES. IMPROPER HANDLING BY THE USER MOST LIKELY CAUSED DAMAGE TO THE ROLLER AND BOTH CUTTER GEARS, WHICH MOST LIKELY CAUSED THE CUSTOMER REPORTED EVENT. THE DEVICE WAS SERVICED AND RETURNED TO THE CUSTOMER.
IT WAS REPORTED THAT THE GEARS WERE DAMAGED ON THE ZIMMER SKIN GRAFT MESHER AND THE DEVICE WOULD NOT MESH PROPERLY. IT WAS ALSO REPORTED THAT IT WAS HARD TO CRANK. THE FACILITY REPORTING THE EVENT IS A CADAVER TISSUE BANK. AS SUCH, THERE WAS NO REPORT OF PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 98106 | ZIMMER SKIN GRAFT MESHER | ZIMMER SKIN GRAFT MESHER | GFD | ZIMMER SURGICAL | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00-7703-015-00, SN#(B)(4),| ZIMMER SKIN GRAFT MESHER CUTTERS:| 00-7703-020-00, SN#(B)(4). |