FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3894149 · Received February 6, 2014

Report

Report Number
8020893-2014-00304
Event Type
Malfunction
Date Received
February 6, 2014
Date of Event
January 1, 2014
Report Date
January 27, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE BREATH DELIVERY UNIT (BDU) PRINTED CIRCUIT BOARD (PCB) AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.

Description of Event or Problem · 1

COVIDIEN RECEIVED INFORMATION STATING THAT AN 840 VENTILATOR WOULD NOT TURN ON. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
78803 840 VENTILATOR CBK; VENTILATOR, CONTINUOUS, FACILITY USE CBK COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND 840

Patients

Seq Age Sex Outcome Treatment
1