FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3894093 · Received June 25, 2014

Report

Report Number
2531779-2014-18239
Event Type
Malfunction
Date Received
June 25, 2014
Report Date
June 22, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE USER¿S SETTINGS SHOWED THAT THE INSULIN ON BOARD (IOB) DURATION WAS SET TO SIX HOURS. DURING TESTING, AN EXERCISE WAS PERFORMED WITH 10 UNIT BOLUS AND A DURATION PERIOD OF 2.0 HOURS. AFTER THE 2 HOURS THE IOB STATUS WAS 0.00 AND THIS WAS ACCURATELY REFLECTED IN THE IOB STATUS. THE INSULIN ON BOARD FEATURE WAS FOUND TO BE FUNCTIONING PROPERLY AND THE COMPLAINT WAS NOT DUPLICATED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS (INSULIN ON BOARD CALCULATION) ISSUE. THE INSULIN ON BOARD DURATION WAS REPORTEDLY SET TO SIX HOURS. THE REPORTER ALLEGED THAT BEFORE BOLUSES FOR THE PAST FEW MONTHS THE INSULIN ON BOARD WOULD BE AT ZERO, EVEN WHEN IT HAD NOT BEEN SIX HOURS SINCE THE LAST BOLUS. THE REPORTER COULD NOT CONFIRM IF THIS OCCURRED WITH EVERY BOLUS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
371974 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 80 YR