LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI)
Report
- Report Number
- 2023950-2014-00010
- Event Type
- Injury
- Date Received
- June 12, 2014
- Report Date
- June 12, 2014
- Manufacturer
- ZEST ANCHORS, LLC
- Product Code
- DZE
- PMA / PMN Number
- K120198
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- DENTIST
Narratives
USER DOCUMENTATION (TECHNIQUE MANUAL, P/N L8019-TM) SPECIFIES THAT IMMEDIATE LOADING IS SUITABLE ONLY IF SUFFICIENT PRIMARY STABILITY OF THE IMPLANT IS ACHIEVED AT THE TIME OF PLACEMENT. ALSO, THE RECOMMENDED IMPLANT INSERTION TORQUE AND IMPLANT ATTACHMENT TORQUE IS 30N-CM. IF THE IMPLANT PLACEMENT TORQUE IS LESS THAN 30 N-CM, THEN THE ATTACHMENTS SHOULD ONLY BE HAND TIGHTENED. THE IMPLANT INSERTION TORQUE SHOULD NOT EXCEED 70 N-CM; IF 70 N-CM IS REACHED PRIOR TO FULL SEATING, THE IMPLANT SHOULD BE REMOVED AND THE OSTEOTOMY SHOULD BE ENLARGED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CLINICIAN: FOUR IMPLANTS WERE PLACED AND WERE IMMEDIATELY LOADED, THREE OF THE FOUR IMPLANTS FAILED TO INTEGRATE. THE CLINICIAN DID NOT SPECIFY THE AMOUNT OF TORQUE USED TO PLACE THE ABUTMENT AND THE IMPLANT. ALSO, THEY DID NOT INDICATE PATIENT'S BONE DENSITY. FAILURE TO OSSEOINTEGRATE IS A WELL-DOCUMENTED INHERENT RISK OF DENTAL IMPLANTS. IN THE MAJORITY OF CASES WHERE AN IMPLANT FAILS TO INTEGRATE WITH THE BONE AND IS REJECTED BY THE BODY, THE CAUSE IS UNKNOWN. THE IMPLANTS' LOT HISTORY RECORDS WERE REVIEWED; ANY DISCREPANCIES OR ISSUES OF NON-CONFORMANCE WERE SUCCESSFULLY RESOLVED PRIOR TO THE RELEASE OF THE PARTS. IMPLANTS WERE MANUFACTURED TO SPECIFICATIONS. NO FURTHER ACTION IS REQUIRED.
IMPLANTS FAILED TO OSSEOINTEGRATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 348893 | LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI) | ENDOSSEIOUS DENTAL IMPLANT | DZE | ZEST ANCHORS, LLC | I0L0K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |