ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-04360
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Date of Event
- May 8, 2014
- Report Date
- May 8, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH TISSUE PAD LIFTED AT THE TIP OF THE CLAMP ARM. UPON VISUAL INSPECTION, IT WAS NOTED THAT THE SHAFT OF THE DEVICE WAS SLIGHTLY BENT. BASED ON THE CONDITION OF THE TISSUE PAD, A POSSIBLE CAUSE FOR THIS DAMAGE IS THAT THE TISSUE PAD GOT STUCK ON A TROCAR OR BODY TISSUE; HOWEVER NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. CARE SHOULD BE TAKEN WHEN INTRODUCING THE INSTRUMENT INTO A TROCAR OR DURING USE INSIDE THE PATIENT. CLEANING OF THE PAD, NOT IN ACCORDANCE WITH THE IFU, CAN ALSO RESULT IN REMOVAL OF THE PAD DURING USE. HOWEVER, NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE FOUND DAMAGE.
(B)(4).
IT WAS REPORTED THAT DURING A LAPAROSCOPIC NEPHRECTOMY PROCEDURE, THE WHITE TISSUE PAD WAS MELTED BUT STILL IN THE CLAMP ARM. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED. ONE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 371129 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |