CATALYS PRECISION LASER SYSTEM
Report
- Report Number
- 3005675890-2014-00021
- Event Type
- Injury
- Date Received
- June 12, 2014
- Date of Event
- May 15, 2014
- Report Date
- June 12, 2014
- Manufacturer
- OPTIMEDICA CORPORATION
- Product Code
- OOE
- PMA / PMN Number
- K121091
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION OF THIS INCIDENT INCLUDED ANALYSIS OF THE SYSTEM DATABASE AND THE SYSTEM OPTICAL COHERENCE TOMOGRAPHY (OCT) RECORDING. THE SYSTEM VIDEO DISPLAY RECORDING AND THE OPERATING ROOM SURGICAL VIDEO WERE NOT AVAILABLE FOR ANALYSIS. FROM THE ANALYSIS OF THE SYSTEM DATABASE AND THE SYSTEM OCT RECORDING IT WAS DETERMINED THAT ALL SETTINGS AND PARAMETERS OF THE SYSTEM WERE FOUND TO BE WITHIN ACCEPTABLE LIMITS. THE SYSTEM DATABASE VIDEO IMAGES INDICATED THE ABSENCE OF, OR MINIMAL, EYE MOVEMENT DURING THE LASER TREATMENT. THE SURGEON REPORTED AND THE SYSTEM DATABASE CONFIRMED THAT A LOSS OF SUCTION DID OCCUR DURING THE FRAGMENTATION PORTION OF THE LASER TREATMENT. ACCORDING TO THE SURGEON, THE ANTERIOR CAPSULOTOMY WAS 100% COMPLETE AND STANDARD CONTINUOUS CURVILINEAR CAPSULORHEXIS (CCC) SURGICAL TECHNIQUE WAS USED TO REMOVE THE CAPSULOTOMY DISC. THE SLIVER WAS NOT NOTICED UNTIL AFTER SEPARATED WITH FORCEPS. THE CAUSE(S) OF THE ANTERIOR CAPSULE TEAR IS UNKNOWN.
IT WAS REPORTED THAT A PATIENT WHO UNDERWENT ANTERIOR CAPSULOTOMY, LENS FRAGMENTATION, AND CORNEAL INCISIONS WITH THE CATALYST SYSTEM (SYSTEM) SUBSEQUENTLY EXPERIENCED AN ANTERIOR CAPSULE TEAR IN THE OPERATING ROOM (OR) DURING THE SURGICAL PROCEDURE TO REMOVE THE CATARACT. THE SURGEON OBSERVED THE CAPSULE TEAR DURING THE REMOVAL OF THE CAPSULOTOMY DISC. NO ADD'L COMPLICATION AND/OR MEDICAL INTERVENTION WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 348668 | CATALYS PRECISION LASER SYSTEM | OPHTHALMIC FEMTOSECOND LASER | OOE | OPTIMEDICA CORPORATION | CATALYS-U |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Other |