CS300
Report
- Report Number
- 2249723-2014-00713
- Event Type
- Death
- Date Received
- June 11, 2014
- Date of Event
- May 12, 2014
- Report Date
- May 12, 2014
- Manufacturer
- DATASCOPE CORP.,
- Product Code
- DSP
- PMA / PMN Number
- K063525
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE COMPANY REP VISITED THE FACILITY AND WAS UNABLE TO DUPLICATE THE COMPLAINT. IT WAS DISCOVERED THAT THE TOP COVER OF THE UNIT WAS CRACKED. THE COMPANY REP STATED ON (B)(4) 2014 THAT THIS CRACKED TOP COVER WAS NOT RELATED TO THE DISPLAY. THE TOP COVER CRACKED WHEN THE CUSTOMER ACCIDENTALLY LOWERED A TABLE ONTO THE CONSOLE. THE COMPANY REP REPLACED THE TOP COVER (PART 0380-00-0497), TOP COVER INSULATOR (PART 0349-00-0348), AND THE LABEL (PART 0334-00-1635-01). THE CRACKED TOP COVER WAS DISPOSED OF BY THE CUSTOMER. THE IABP WAS TESTED TO FACTORY SPECIFICATIONS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO THE CUSTOMER. THE PRODUCTION DEVICE RECORD FOR THE IABP INVOLVED IN THE EVENT WAS REVIEWED. THERE ARE NO NON-CONFORMANCES RELATED TO THE REPORTED EVENT. (B)(4).
THE PT WAS 5' 7". THE BALLOON THERAPY BEGAN ON SATURDAY (B)(6) 2014. THE CUSTOMER STATED THEY WOULD LOSE THE PRESSURE WAVEFORM. THEY WERE ABLE TO FLUSH THE LUMEN A FEW TIMES AND WERE ABLE TO ZERO AND RE-ESTABLISH THE PRESSURE WAVEFORM. THE PROBLEM OCCURRED A FEW TIMES BETWEEN SATURDAY ((B)(6) 2014). THE CUSTOMER SUPPLIED STRIPS SHOWING DAMPENING AND LOSS OF WAVEFORMS. AFTER THE DEVICE LOST ALL PT WAVEFORMS, THE DISPLAY WENT BLANK. THE THERAPY WAS NEVER INTERRUPTED. THE CS300 CONTINUED TO PUMP. AT APPROXIMATELY 4:00PM ON MONDAY ((B)(6) 2014), THEY SWITCHED CONSOLES. THE CUSTOMER DID NOT HAVE ANY MORE WAVEFORM ISSUES. THE PT EXPIRED ON MONDAY EVENING AT 10PM. THE PT WAS VERY SICK AND UNSTABLE. THE CUSTOMER IS NOT ATTRIBUTING THE DEATH TO THE BALLOON PUMP. CONTACT NURSE MGR ICU (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 344900 | CS300 | INTRA-AORTIC BALLOON PUMP | DSP | DATASCOPE CORP., | CS300 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Death |