XR2 METAL-FREE CONVERSION ADAPTOR
Report
- Report Number
- 3004608878-2014-00085
- Event Type
- Injury
- Date Received
- June 9, 2014
- Date of Event
- May 19, 2014
- Report Date
- May 19, 2014
- Manufacturer
- INTEGRA LIFESCIENCES CORPORATION OH/USA
- Product Code
- HBL
- PMA / PMN Number
- K130389
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
TO DATE, THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED UPON THE REPORTED INFORMATION.
IT WAS REPORTED THAT WHEN ATTEMPTING TO REMOVE THE SKULL CLAMP FROM THE PATIENT AFTER THE SURGERY, THE METAL FREE CONVERSION ATTACHMENT WOULD NOT RELEASE. AFTER SEVERAL ATTEMPTS, A SURGICAL INSTRUMENT WAS USED TO RELEASE THE PISTON HOLDING THE PIN. THE SAPPHIRE PIN THAT WAS USED IN THE SURGERY CRACKED AND THE TIP BROKE OFF. THE VERY TIP OF THE PIN COULD STILL BE IN THE PATIENT SKULL. THE SURGEON AND STAFF COULD NOT VERIFY. ADDITIONAL INFORMATION WAS REQUESTED AND ON (B)(6) 2014, THE FOLLOWING WAS REPORTED BY THE SALES REPRESENTATIVE: ON (B)(6) 2014, A (B)(6) YEAR OLD MALE PATIENT UNDERWENT A CRANIOTOMY FOR TUMOR REMOVAL. SURGICAL PLIER WAS THE SURGICAL INSTRUMENT USED TO RELEASE THE PISTON HOLDING THE PIN. THE THREE ADULT SAPPHIRE PINS USED IN THE SURGERY WERE THE A2020 (MAYFIELD RADIOLUCENT SKULL PINS). LOT NUMBER OF THE PINS WERE NOT KNOWN. IT WAS CONFIRMED THAT THE TIP OF THE PIN WAS LIKELY STILL IN THE PATIENT SKULL. NO PATIENT TREATMENT WAS DONE. IT WAS REPORTED THAT MOST LIKELY, THE SURGEON WILL REMOVE THE PIN IF IT DOES NOT WORK ITSELF OUT EVENTUALLY OR IF IT CAUSES ADDITIONAL PROBLEMS. IF THEY STILL HAVE IT, THE BROKEN PIN WILL POTENTIALLY BE RETURNED TO INTEGRA FOR EVALUATION. THE XR2 METAL FREE CONVERSION ADAPTOR WILL BE SENT BACK TO INTEGRA NEXT WEEK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 334302 | XR2 METAL-FREE CONVERSION ADAPTOR | NEURO | HBL | INTEGRA LIFESCIENCES CORPORATION OH/USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Other | A2020 (MAYFIELD RADIOLUCENT SKULL PINS) |