FDA Adverse Event
Malfunction
Summary report: N
ADVANTA
MDR report key: 3893553
·
Received February 15, 2014
Report
- Report Number
- 1824206-2014-00452
- Event Type
- Malfunction
- Date Received
- February 15, 2014
- Date of Event
- January 17, 2014
- Report Date
- January 17, 2014
- Manufacturer
- HILL-ROM INC
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
HILL-ROM SERVICE TECHNICIAN INSPECTED THE BED AND FOUND THAT THE CAUSE OF THE SIDE RAIL NOT LATCHING WAS A BENT SLIDE BRACKET. THE BRACKET WAS REPLACED AND THE BED FUNCTIONED AS EXPECTED. THE MOST PROBABLE CAUSE FOR THE BENT SLIDE BRACKET IS MECHANICAL DAMAGE. NO FURTHER ACTION REQUIRED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWS THAT THE LAST PREVENTIVE MAINTENANCE WAS AUGUST 2013.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING PT RIGHT HEAD SIDE RAIL STUCK IN DOWN POSITION. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 99591 | ADVANTA | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC | P1600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |