FDA Adverse Event Malfunction Summary report: N

ADVANCE

MDR report key: 3893540 · Received February 14, 2014

Report

Report Number
1824206-2014-00438
Event Type
Malfunction
Date Received
February 14, 2014
Date of Event
January 16, 2014
Report Date
January 16, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

HILL-ROM SERVICE TECHNICIAN INSPECTED THE BED AND FOUND THAT BOTH BRAKE AND STEER CASTERS WERE NOT WORKING PROPERLY. HE REPLACED THE BRAKE AND STEER CASTERS AND THE BED FUNCTIONED AS EXPECTED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWS THAT THE LAST PREVENTIVE MAINTENANCE PERFORMED BY HILL-ROM WAS JANUARY 2012. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM A USER FACILITY STATING, "BRAKES NOT WORKING PROPERLY, BED STILL ROLLS WHEN BRAKE IS SET. THERE WAS NO PATIENT/INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
99297 ADVANCE A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. P1165

Patients

Seq Age Sex Outcome Treatment
1