FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR

MDR report key: 3892863 · Received June 24, 2014

Report

Report Number
3015876-2014-00699
Event Type
Malfunction
Date Received
June 24, 2014
Date of Event
May 15, 2014
Report Date
May 29, 2014
Manufacturer
PHYSIO-CONTROL, INC
Product Code
LDD
PMA / PMN Number
K063119
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EXAMINED THE CUSTOMER'S DEVICE AND VERIFIED THE REPORTED ISSUE. PHYSIO THEN REPLACED BOTH THE SYSTEM CONTROLLER (SC) PCB AND USER INTERFACE (UI) PCB ASSEMBLIES AND AFTER OBSERVING PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING THE UNIT WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EXAMINED THE REMOVED SC PCB ASSEMBLY WHERE IT WAS DETERMINED THAT THE CAUSE OF THE REPORTED ISSUE WAS AN INTEGRATED CIRCUIT (IC) CHIP, DESIGNATOR U61. THE IC CHIP PUT THE UNIT INTO A CONSTANT RESET CONDITION, WHICH WOULD NOT ALLOW THE DEVICE TO COMPLETE THE BOOT UP CYCLE AND PREVENTED DEFIBRILLATION FROM BEING DELIVERED. THE REMOVED UI PCB ASSEMBLY WAS AN ANCILLARY PART AND THERE WAS NO FAILURE OF THIS ASSEMBLY.

Description of Event or Problem · 1

THE CUSTOMER, A BIOMEDICAL ENGINEER, CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE HAD A SERVICE INDICATOR PRESENT. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE DEVICE, PHYSIO-CONTROL OBSERVED THAT IT WOULD NOT COMPLETE THE BOOT UP CYCLE AND, AS A RESULT, WOULD NOT BE ABLE TO PROVIDE THERAPY, IF NECESSARY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
369249 LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) LDD PHYSIO-CONTROL, INC 20

Patients

Seq Age Sex Outcome Treatment
1