FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA CRT-D, DF-4 CONNECTOR

MDR report key: 3892723 · Received June 24, 2014

Report

Report Number
2938836-2014-12398
Event Type
Malfunction
Date Received
June 24, 2014
Date of Event
May 15, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED IN-CLINIC FOR FOLLOW UP. THE DEVICE WAS EXHIBITING POST-PACED T-WAVE OVERSENSING ON THE VENTRICULAR CHANNEL VIA REVIEW OF REAL-TIME EGM. THE ICD WAS REPROGRAMMED AND REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
368682 QUADRA ASSURA CRT-D, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3365-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 47 YR