FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3892705 · Received June 24, 2014

Report

Report Number
1416980-2014-20165
Event Type
Malfunction
Date Received
June 24, 2014
Date of Event
April 15, 2014
Report Date
May 30, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED FOR THE REPORTED EVENT. A REVIEW OF THE EVENT HISTORY LOG VERIFIED THE OCCURRENCE OF AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE THAT ONE OR MORE CYCLES ADVANCED TO THE NEXT FILL WHEN SLOW/NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2014 AT 00:41:53. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 1827ML, INDICATING THE HOME PATIENT (HP) DRAINED 1827ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2400ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
368674 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1