FDA Adverse Event
Malfunction
Summary report: N
V60 VENTILATOR
MDR report key: 3892640
·
Received June 24, 2014
Report
- Report Number
- 2031642-2014-00546
- Event Type
- Malfunction
- Date Received
- June 24, 2014
- Report Date
- May 27, 2014
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- MNT
- PMA / PMN Number
- K082660
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
OUT OF WARRANTY REPAIR.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT WHEN THE UNIT WAS TURNED ON. THE UNIT KEEPS BEEPING AND THE SCREEN IS BLANK. THERE WAS NO PATIENT INVOLVEMENT. THE CUSTOMER SPOKE TO TECH SUPPORT AND ADVISED HIM TO UNPLUGGED THE BATTERY AND TRY A DIFFERENT BATTERY AND IF THE UNIT STILL BEEPING AND NOT POWERING UP, TECH SUPPORT SAID THAT IT COULD BE THE PM BOARD. THE CUSTOMER UNPLUGGED THE BATTERY AND RECONNECTED IT BACKUP AND THE UNIT STOPS BEEPING AND THE UNIT POWERED UP. THE CUSTOMER SAID THAT NO PARTS WAS REPLACED AND THE UNIT IS NOW WORKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 369242 | V60 VENTILATOR | VENTILATOR, CONTINUOUS | MNT | RESPIRONICS CALIFORNIA, INC. | V60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |