PK DISSECTING FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-03872
- Event Type
- Malfunction
- Date Received
- June 24, 2014
- Date of Event
- May 22, 2014
- Report Date
- May 28, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS FOUND THAT THE CABLE WAS NOT ONLY FRAYED, BUT COMPLETELY BROKEN. THE PITCH UP CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE DAMAGED CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT THE PK DISSECTING FORCEPS INSTRUMENT WIRE WAS NOTED TO BE FRAYED DOWN AT THE TIP DURING REPROCESSING. NO FRAGMENTS FELL INTO THE PATIENT, NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.