I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 3892360 · Received June 24, 2014

Report

Report Number
2245578-2014-00027
Event Type
Malfunction
Date Received
June 24, 2014
Date of Event
June 11, 2014
Report Date
June 11, 2014
Manufacturer
ABBOTT POINT OF CARE
Product Code
MMI
PMA / PMN Number
K031739
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION WAS COMPLETED ON 07/17/2014. RETAIN AND RETURNED CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

ON (B)(6) 2014, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT OF 0.41 ON PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. DATE TIME METHOD RESULT SAMPLE(B)(6) 2014 09:35 I-STAT 0.41 A(B)(6) 2014 09:50 I-STAT 0.12 A(B)(6) 2014 10:22 I-STAT 0.00 B(B)(6) 2014 10:22 LAB 0.00 BTHERE WERE NO INJURIES REPORTED WITH THIS EVENT. AN INVESTIGATION HAS BEEN INITIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
367877 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE NA P14073A

Patients

Seq Age Sex Outcome Treatment
1