I-STAT TROPONIN (CTNI) CARTRIDGE
Report
- Report Number
- 2245578-2014-00027
- Event Type
- Malfunction
- Date Received
- June 24, 2014
- Date of Event
- June 11, 2014
- Report Date
- June 11, 2014
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- MMI
- PMA / PMN Number
- K031739
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). THE INVESTIGATION WAS COMPLETED ON 07/17/2014. RETAIN AND RETURNED CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.
(B)(4).
ON (B)(6) 2014, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT OF 0.41 ON PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. DATE TIME METHOD RESULT SAMPLE(B)(6) 2014 09:35 I-STAT 0.41 A(B)(6) 2014 09:50 I-STAT 0.12 A(B)(6) 2014 10:22 I-STAT 0.00 B(B)(6) 2014 10:22 LAB 0.00 BTHERE WERE NO INJURIES REPORTED WITH THIS EVENT. AN INVESTIGATION HAS BEEN INITIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 367877 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE | NA | P14073A |