FDA Adverse Event Malfunction Summary report: N

ARMADA 14 PTA CATHETER

MDR report key: 3891989 · Received June 24, 2014

Report

Report Number
2024168-2014-04040
Event Type
Malfunction
Date Received
June 24, 2014
Date of Event
May 28, 2014
Report Date
May 30, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
LIT
PMA / PMN Number
K102705
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION AND THE REPORTED DIFFICULTY INFLATING WAS NOT CONFIRMED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES. A QUERY OF THE COMPLAINT HANDLING DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT USING A FEMORAL ARTERY ACCESS APPROACH DURING A PROCEDURE OF THE LOWER EXTREMITY, THE 2.0 X 200 MM ARMADA 14 BALLOON DILATATION CATHETER (BDC) WAS POSITIONED BUT COULD NOT BE INFLATED AT THE LESION. IT WAS NOTED THAT PLACEMENT WAS INITIATED WITHOUT ANY ISSUE; A MANOMETER WAS USED BUT THE BALLOON DID NOT INFLATE. A SECOND ARMADA BDC WAS USED SUCCESSFULLY WITHOUT REPORTED ISSUE. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
368865 ARMADA 14 PTA CATHETER PERIPHERAL DILATATION CATHETER LIT AV-TEMECULA-CT 4010841

Patients

Seq Age Sex Outcome Treatment
1 SHEATH: CORDIS BRITE TIP 4F