TRANSTAR STRETCHER
Report
- Report Number
- 3006697241-2014-00105
- Event Type
- Malfunction
- Date Received
- February 4, 2014
- Date of Event
- January 8, 2014
- Report Date
- January 8, 2014
- Manufacturer
- HILLROM DE MEXICO S DE RL DE CV
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- UNKNOWN
Narratives
THE TECHNICIAN FOUND THE BRAKE CASTERS ARE WORN AND THE BRAKE CASTERS WOULD ROTATE IN A CIRCLE IF THE STRETCHER WAS PUSHED ON THE SIDE WHILE IN BRAKE MODE. THE DEVICE REFERENCED IN THIS REPORT IS OVER 12 MONTHS OLD AND ACCORDING THE RECOMMENDED PREVENTATIVE MAINTENANCE SCHEDULE AT LEAST ONE PM SHOULD BE PERFORMED EVERY 12 MONTHS. THEREFORE THE ACCEPTANCE CRITERIA, BASED ON THE DHR, WERE DETERMINED TO BE ACCEPTABLE AT THE TIME THE DEVICE WAS RELEASED. GIVEN THE TIME WHICH HAS ELAPSED, THE ALLEGED FAILURE OF THE DEVICE WOULD NOT BE RELATED TO ANY ISSUE WHICH OCCURRED DURING THE MANUFACTURING PROCESS. THE TECHNICIAN REPLACED THE BRAKE CASTERS TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
THE ACCOUNT REPORTED THE BRAKE CASTERS WOULD ROTATE IN A CIRCLE WHILE IN BRAKE MODE. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).