FDA Adverse Event Malfunction Summary report: N

LUMENIS ONE

MDR report key: 3891595 · Received June 24, 2014

Report

Report Number
3004135191-2014-00045
Event Type
Malfunction
Date Received
June 24, 2014
Date of Event
June 18, 2014
Report Date
June 24, 2014
Manufacturer
LUMENIS, LTD.
Product Code
GEX
PMA / PMN Number
K060448
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

LUMENIS INVESTIGATED THE REPORTED COMPLAINT. A LUMENIS TECHNICAL EXPERT COMPLETED AN EXAMINATION OF THE SUBJECT DEVICE IDENTIFYING A CRACK IN A COOLANT TUBE FITTING TO BE THE CAUSE OF THE EVENT REPORT. LUMENIS IS REPORTING THIS COMPLAINT SINCE IT IS SIMILAR TO AN EVENT IN WHICH THE COOLING SYSTEM LEAK RESULTED IN COMBUSTION OF THE HIGH VOLTAGE POWER SUPPLY.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT A COOLING SYSTEM LEAK OCCURRED IN A LUMENIS ONE LASER. NO REPORT OF RELATED INJURY WAS RECEIVED.