HAKIM INLINE PROGRAMMABLE VALVE SG
Report
- Report Number
- 1226348-2014-11738
- Event Type
- Injury
- Date Received
- June 24, 2014
- Date of Event
- May 23, 2014
- Manufacturer
- CODMAN & SHURTLEFF
- Product Code
- JXG
- PMA / PMN Number
- PK992173
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE VALVE WAS RETURNED FOR EVALUATION. THE INVESTIGATION OF THE RETURNED VALVE DID CONFIRM A PROBLEM. THE LOT NUMBER WAS CPPBVT, SERIAL NUMBER (B)(4), PRODUCT CODE 82-3162. THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS 50MMH2O. THE VALVE WAS VISUALLY INSPECTED, NO DEFECTS WERE NOTED. THE VALVE WAS IRRIGATED WITH PURIFIED WATER; AN OCCLUSION WAS NOTED AT THE INLET OF THE RUBY BALL. THEREFORE A FINE NEEDLE WAS INSERTED INTO THE FLOW OPENING OF THE VALVE INLET UNDER THE RUBY BALL AND ITS SEAT, THE RUBY BALL WAS MANUALLY POPPED FREE FROM ITS STUCK POSITION USING LIGHT FORCE. THE VALVE WAS IRRIGATED AGAIN, NO OCCLUSION WAS NOTED. THE SIPHON GUARD WAS IRRIGATED WITH PURIFIED WATER, NO OCCLUSION WAS NOTED. THE VALVE WAS DRIED. THE VALVE WAS LEAK TESTED AND NO LEAKS WERE NOTED. THE VALVE WAS REFLUX TESTED, THE VALVE PASSED THE TEST. THE VALVE WAS TESTED FOR PROGRAMMING AND THE VALVE FAILED THE TEST, DURING THE PROGRAMMING PROCESS THE CAM MECHANISM DID NOT MOVE. THE VALVE WAS THEN PRESSURE TESTED AT 50MMH2O, THE VALVE PASSED THE TEST. THE VALVE WAS DISMANTLED AND WAS EXAMINED UNDER MICROSCOPE AT APPROPRIATE MAGNIFICATION: BIOLOGICAL DEBRIS WAS FOUND ON THE SPRING, ON THE SPRING PILLAR, ON THE RUBY BALL, ON THE SEAT OF THE RUBY BALL, ON THE CAM MECHANISM, ON THE CAM MECHANISM PILLAR, AND ON THE BASE PLATE. REVIEW OF THE HISTORY DEVICE RECORDS CONFIRMED THE VALVE, PRODUCT CODE 82-3162 WITH LOT NUMBER CPPBVT, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK ON THE 28TH MARCH 2014. THE ROOT CAUSE OF THE OCCLUSION AND PROGRAMMING PROBLEM FOUND DURING INVESTIGATION IS DUE TO BIOLOGICAL DEBRIS FOUND ON THE SPRING, ON THE SPRING PILLAR, ON THE RUBY BALL, ON THE SEAT OF THE RUBY BALL, ON THE CAM MECHANISM, ON THE CAM MECHANISM PILLAR, AND ON THE BASE PLATE. NO CORRECTIVE ACTION IS NEEDED BASED ON THE RESULTS OF THE EVALUATION. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CONSIDERED CLOSED.
UPON COMPLETION OF INVESTIGATION, A FOLLOW UP REPORT WILL BE FILED. DEVICE IS UNDER INVESTIGATION.
NO REPLACEMENT NEEDED. VALVE SEEMED TO FLOW QUICKER THEN NORMAL IN THE 10-15MM RANGE ON THE MANOMETER TEST. MD WOULD LIKE VALVE TESTED AND BE NOTIFIED OF THE RESULTS. (B)(6) 2014 PER THE REPS DIRECTION REPLACEMENT IS NOT NEEDED. REP ALSO CONFIRMED THAT THE DEVICE WAS REVISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 368411 | HAKIM INLINE PROGRAMMABLE VALVE SG | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF | CPPBVT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |