ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-18103
- Event Type
- Malfunction
- Date Received
- June 24, 2014
- Report Date
- June 20, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: DURING TESTING, EVIDENCE OF MOISTURE IN THE PUMP. THE PUMP FAILED A LEAK TEST DUE TO CRACKED CASING. THE DISPLAY SCREEN WAS DIM AND DISCOLORED. THE DISPLAY SCREEN WAS SCRATCHED WHICH HAS NO INFLUENCE ON INSULIN DELIVERY FUNCTION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED MOISTURE IN THE PUMP AND A DISPLAY FAILURE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 368720 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |