FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3891407 · Received June 24, 2014

Report

Report Number
1416980-2014-20065
Event Type
Malfunction
Date Received
June 24, 2014
Date of Event
December 26, 2013
Report Date
May 30, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. REVIEW OF THE EVENT LOG FOUND EVIDENCE OF THE REPORTED INCREASED INTRAPERITONEAL VOLUME (IIPV) CONDITION. ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. THE CAUSE WAS UNABLE TO BE DETERMINED WITH THE PROVIDED INFORMATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN ERROR 105 (NIGHT DRAIN #5) ALARM WAS IDENTIFIED IN THE LOG. THIS ALARM INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE, INDICATING AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. THE ALARM OCCURRED ON (B)(6) 2013 AT 11:31:54. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
368438 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 45 YR