SPECTRUM INFUSION PUMP
MDR report key: 3891143
·
Received January 31, 2014
Report
- Report Number
- 1314492-2014-05396
- Event Type
- Malfunction
- Date Received
- January 31, 2014
- Date of Event
- December 1, 2013
- Report Date
- January 6, 2014
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF (B)(4). THE IDENTIFIED "DOOR JAMMED" ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFORMATION IS RECEIVED A FOLLOW UP WILL BE SENT. THE UPPER AUXILIARY SUB-ASSEMBLY, INPUT/OUTPUT PRINTED CIRCUIT BOARD, AND DOOR LATCH HOOK WERE REPLACED. SEE SCANNED PAGE.
Description of Event or Problem · 1
DURING THE EVALUATION OF A RETURNED SPECTRUM INFUSION PUMP, 5 OCCURRENCES OF "DOOR JAMMED¿ ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNKNOWN SINCE THESE ITEMS WERE FOUND DURING EVALUATION OF THE DEVICE.