FDA Adverse Event Malfunction Summary report: N

TAXUS? LIBERTÉ?

MDR report key: 3891119 · Received June 24, 2014

Report

Report Number
2134265-2014-03550
Event Type
Malfunction
Date Received
June 24, 2014
Date of Event
April 18, 2014
Report Date
May 30, 2014
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P060008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS COMBINATION PRODUCT. (B)(4). DEVICE EVALUATED BY MFR: A VISUAL EXAMINATION OF THE CRIMPED STENT FOUND THE STENT WAS DAMAGED. THE STENT STRUTS AT THE PROXIMAL END OF THE STENT WERE BUNCHED AND DISTORTED. A SNAP GAUGE WAS USED TO MEASURE THE CRIMPED STENT OD AT THE DISTAL END. THE SPECIFIED STENT PROTECTOR WAS NOT TOO TIGHT ON THE CRIMPED STENT PROVIDED THE CORRECT REMOVAL OF THE STENT PROTECTOR WAS USED. THE BALLOON CONES PROFILES WERE REVIEWED AND NO ISSUES WERE NOTED WITH THE OVERALL BALLOON PROFILE. THE BALLOON WINGS WERE TIGHTLY WRAPPED AND EVENLY FOLDED AND WERE NOT SUBJECTED TO POSITIVE PRESSURE. A VISUAL AND TACTILE EXAMINATION OF THE HYPOTUBE AND THE SHAFT POLYMER EXTRUSION FOUND NO ISSUES WITH THE PROFILE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON ANALYSIS APPROVED ON (B)(4) 2014. IT WAS REPORTED THAT CROSSING DIFFICULTIES OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT CIRCUMFLEX (LCX) ARTERY. A 12 X 2.25MM TAXUS¿ LIBERTÉ¿ STENT WAS SELECTED TO TREAT THE LESION; HOWEVER, THE DEVICE WAS UNABLE TO CROSS THE TARGET LESION. HOWEVER, DEVICE ANALYSIS REVEALED STENT DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
367240 TAXUS? LIBERTÉ? CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - GALWAY H7493894012220 16120549

Patients

Seq Age Sex Outcome Treatment
1 58 YR