SPECTRUM INFUSION PUMP

MDR report key: 3891035 · Received January 31, 2014

Report

Report Number
1314492-2014-05319
Event Type
Malfunction
Date Received
January 31, 2014
Date of Event
January 1, 2014
Report Date
January 6, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED (THROUGH THE HISTORY LOG) BUT DID NOT REPRODUCE THE SYSTEM ERROR 322. THE DEVICE WAS OPERATING OUT OF SPECIFICATION. PAST REPAIRS HAVE SHOWN THAT THE UPPER AND LOWER AUXILIARIES TO BE THE FAILING COMPONENTS. THE UPPER AND LOWER AUXILIARIES WERE REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP ALARMED SYSTEM ERROR 322. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.