FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX?

MDR report key: 3890976 · Received June 24, 2014

Report

Report Number
2134265-2014-03679
Event Type
Malfunction
Date Received
June 24, 2014
Date of Event
May 30, 2014
Report Date
June 2, 2014
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
K121667
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: RETURNED PRODUCT CONSISTED OF AN NC QUANTUM APEX BALLOON CATHETER. THERE WAS BLOOD IN THE INFLATION LUMEN AND GUIDEWIRE LUMEN. THE DEVICE WAS RETURNED IN TWO (2) PIECES. THE CATHETER WAS VISUALLY AND TACTILELY INSPECTED. THERE WERE MULTIPLE HYPOTUBE KINKS AND THE HYPOTUBE WAS BROKEN 83CM FROM THE STRAIN RELIEF. THE FRACTURE FACES WERE OVAL SHAPED, AS IF KINKED PRIOR TO SEPARATION. THE SHAFT WAS KINKED 26.5CM FROM THE DISTAL TIP. THE CATHETER WAS INSPECTED UNDER MAGNIFICATION; THE DISTAL TIP WAS DAMAGED BUT THERE WAS NO OTHER DAMAGE OR IRREGULARITIES. THERE WAS NO EVIDENCE OF ANY MATERIAL OR MANUFACTURING DEFICIENCIES CONTRIBUTING TO THE DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SHAFT BREAK OCCURED. A 15MM X 4.00MM NC QUANTUM APEX¿ BALLOON CATHETER WAS SELECTED FOR THE PROCEDURE, HOWEVER THE SHAFT OF THE CATHETER WAS BROKEN.

Description of Event or Problem · 1

IT WAS REPORTED THAT SHAFT BREAK OCCURED. A 15MM X 4.00MM NC QUANTUM APEX BALLOON CATHETER WAS SELECTED FOR THE PROCEDURE, HOWEVER THE SHAFT OF THE CATHETER WAS BROKEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
367466 NC QUANTUM APEX? CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912415400 16726298