FDA Adverse Event Malfunction Summary report: N

ADVIA CENTAUR XP PSA ASSAY

MDR report key: 3890634 · Received June 24, 2014

Report

Report Number
1219913-2014-00149
Event Type
Malfunction
Date Received
June 24, 2014
Date of Event
May 23, 2014
Report Date
May 27, 2014
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
Product Code
MTF
PMA / PMN Number
P950021
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CAUSE FOR THE FALSE LOW PSA PATIENT RESULT ON THE ADVIA CENTAUR XP SYSTEM WHEN COMPARED TO THE ELEVATED RESULTS ON TWO ALTERNATE PSA TEST METHODS AT ANOTHER SITE MOST LIKELY IS DUE TO TO A HIGH-DOSE HOOK EFFECT. THE PATIENT SAMPLE WHEN DILUTED AND RUN ON AN ADVIA CENTAUR SYSTEM RESULTED IN AN ELEVATED PSA TEST RESULT. THE INSTRUCTION FOR USE UNDER THE PROCEDURAL NOTES HIGH-DOSE HOOK EFFECT STATES THE FOLLOWING: "PATIENT SAMPLES WITH HIGH TOTAL PSA LEVELS CAN CAUSE A PARADOXICAL DECREASE IN THE RLUS (HIGH-DOSE HOOK EFFECT). IN THIS ASSAY, PATIENT SAMPLES WITH TOTAL PSA LEVELS AS HIGH AS 10,000 NG/ML (10,000 UG/L) WILL ASSAY GREATER THAN 100 NG/ML (100 UG/L). THE INSTRUCTION FOR USE UNDER THE LIMITATION SECTION STATES THE FOLLOWING: "NOTE: DO NOT INTERPRET LEVELS OF PSA AS ABSOLUTE EVIDENCE OF THE PRESENCE OR THE ABSENCE OF MALIGNANT DISEASE. BEFORE TREATMENT, PATIENTS WITH CONFIRMED PROSTATE CARCINOMA FREQUENTLY HAVE LEVELS OF PSA WITHIN THE RANGE OBSERVED IN HEALTHY INDIVIDUALS. ELEVATED LEVELS OF PSA CAN BE OBSERVED IN PATIENTS WITH NONMALIGNANT DISEASES. MEASUREMENTS OF PSA SHOULD ALWAYS BE USED IN CONJUNCTION WITH OTHER DIAGNOSTIC PROCEDURES, INCLUDING INFORMATION FROM THE PATIENT'S CLINICAL EVALUATION." "THE CONCENTRATION OF TOTAL PSA IN A GIVEN SPECIMEN DETERMINED WITH ASSAYS FROM DIFFERENT MANUFACTURERS' CAN VARY DUE TO DIFFERENCES IN ASSAY METHODS, CALIBRATION, AND REAGENT SPECIFICITY. TOTAL PSA DETERMINED WITH DIFFERENT MANUFACTURERS ASSAYS WILL VARY DEPENDING ON THE METHOD OF STANDARDIZATION AND ANTIBODY SPECIFICITY. WARNING: DO NOT PREDICT DISEASE RECURRENCE SOLELY ON SERIAL PSA VALUES." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

A FALSE LOW ADVIA CENTAUR XP PSA RESULT WAS OBTAINED BY THE CUSTOMER ON A PATIENT SAMPLE AND THE RESULT WAS REPORTED TO THE DOCTOR. THE LOW PSA RESULT WAS CONSIDERED DISCORDANT WHEN COMPARED TO ELEVATED RESULTS ON TWO ALTERNATE PSA TEST METHODS AT ANOTHER SITE. THE PATIENT SAMPLE WHEN DILUTED AND REPEATED ON AN ADVIA CENTAUR SYSTEM WAS ELEVATED. THERE WAS NO KNOWN REPORT OF PATIENT TREATMENT BEING ALTERED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT LOW ADVIA CENTAUR XP PSA TEST RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
367692 ADVIA CENTAUR XP PSA ASSAY PSA IMMUNOASSAY, PRODUCT CODE: MTF SIEMENS HEALTHCARE DIAGNOSTICS, INC. N/A 252

Patients

Seq Age Sex Outcome Treatment
1