FDA Adverse Event Other Summary report: N

HELIX-REVOLUTION ACP SYSTEM

MDR report key: 3890139 · Received June 13, 2014

Report

Report Number
2031966-2014-00039
Event Type
Other
Date Received
June 13, 2014
Date of Event
May 14, 2014
Report Date
June 12, 2014
Manufacturer
NUVASIVE, INC.
Product Code
KWQ
PMA / PMN Number
K093804
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

NUVASIVE REF: (B)(4). RADIOGRAPH CONFIRMED THE REPORTED EVENT. RETURNED DEVICES WERE MEASURED AND MET DESIGN REQUIREMENTS. RETURNED DEVICES WERE EVALUATED. REVIEW OF THE DEVICE HISTORY RECORDS NOTES NO MATERIAL NON-CONFORMANCES OR MANUFACTURING ERRORS THAT MAY HAVE CAUSED OR CONTRIBUTED TO THIS MODE OF FAILURE. PARTS MET ACCEPTANCE CRITERIA UPON RELEASE. PT ACTIVITY AT THE TIME OR PRIOR TO THE EVENT IS UNK. PT'S BONE QUALITY IS UNK. IT IS UNK IF THE PT COMPLIED WITH POST-OP CARE INSTRUCTIONS OR SUSTAINED AN IMPACT OF SOME SORT. THE PARTIAL CORPECTOMY ONLY ALLOWED 4 OR THE 8 POINTS OF FIXATION TO BE UTILIZED. THIS LIMED FIXATION MAY BE A CONTRIBUTING FACTOR. THE ANGULATION OF THE SUPERIOR SCREW MAY ALSO BE A FACTOR. THE ROOT CAUSE OF THE ISSUE REMAINS UNK.

Description of Event or Problem · 1

ON (B)(6) 2014, IT WAS REPORTED THAT THE RIGHT SUPERIOR CERVICAL BONE SCREW BACKED OUT OF A THREE LEVEL (C3-C6) ACDF PLATE AT C3. THE PLATE AND SCREWS WERE REMOVED AND REPLACED. REVISION ALSO INCLUDED SUPPLEMENTAL POSTERIOR FIXATION FROM C3-C7. PT IS REPORTEDLY DOING WELL POST REVISION. THE INITIAL SURGERY OCCURRED (B)(6) 2014 IN WHICH SUPPLEMENTAL CERVICAL PLATING WAS PLACED OVER A CERVICAL VERTEBRAL BODY REPLACEMENT DEVICE. THE PARTIAL CORPECTOMY SURGERY OF C4 AND C5 VERTEBRAL BODIES ONLY ALLOWED FOR FIXATION SCREWS TO BE PLACED C3 AND C6 ANTERIORLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350365 HELIX-REVOLUTION ACP SYSTEM SPINAL INT VERT BODY FIXATION APPLIANCE KWQ NUVASIVE, INC. 7801115 M23385

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention